QA-CAL-2026-Q2

SMU calibration & laboratory compliance logbook (ISO/IEC 17025)

Identity

Sample IDQA-CAL-2026-Q2
TitleSMU calibration & laboratory compliance logbook (ISO/IEC 17025)
StatusCompleted
OperatorQuality officer: H. Arslan
Laboratory / FacilityMikroFab Metrology / Quality Unit

Notes

Tagsquality, calibration, ISO 17025, traceability, GLP
Notes2026 Q2 periodic SMU verification + compliance records.

Process Steps

Primary instrument under calibration (Equipment / Instrument)

Instrument NameSource-measure unit (SMU)
ModelKeithley 2636B
Serial No4102217
LocationMeasurement room 1
Calibration Date2027-06-25
SettingsReference: Fluke 8588A (traceable, certificate #FL-2026-0412).
Reservation / UserMetrology — closed

Standards followed (Standard / Compliance)

OrganizationNumberTitleClauseStatus
ISO/IEC17025:2017Competence of testing and calibration laboratories6.4, 7.2Compliant
JCGM100:2008 (GUM)Guide to the expression of uncertainty in measurementApplied
ASTME2655Uncertainty reporting guideReference

Verification test plan (source/measure) (Test Plan / Result)

TestMethod / StandardExpectedResultPass
Voltage source accuracy (10 V)ISO 17025 / GUM±0.02% + 2.4 mV+1.1 mV
Current measurement accuracy (1 mA)Fluke 8588A comparison±0.05% + 100 nA+220 nA
Low-current leakage (1 nA range)short-circuit reading≤50 fA32 fA
Warm-up drift (1 h)10 V constant, logged≤5 ppm3 ppm

Laboratory electrical safety note (Safety / SDS Risk)

Chemical / AgentHigh-voltage SMU (±200 V)
GHS Class / H-Statements
HazardElectric shock; exposed tip at the probe station
PPE (Personal Protective)Insulated gloves not required (<±42 V operation); lockout-tagout at ≥±42 V
Exposure Limit
First AidShock: cut power, 112, CPR-trained personnel.
Risk LevelMedium

Periodic verification procedure (SOP) (Protocol / SOP)

VersionSOP-CAL-03 Rev 2
Scope12-month periodic verification of all SMU/DMM units
Materials / ReagentsTraceable reference (Fluke 8588A), temperature/humidity logging
Procedure (step by step)1) 1 h warm-up. 2) 5-point voltage + 5-point current. 3) Deviation and uncertainty calculation (GUM). 4) Issue certificate. 5) Update label.
HazardsLockout-tagout during high-voltage steps
ReferencesISO 17025:2017 §7.2; GUM

Calibration approval and certificate (Approval / Signature)

Performed byH. Arslan (quality officer)
Reviewed byDr. E. Yılmaz
Approved byProf. A. Kara (lab manager)
Date2026-06-26
Role / TitleQuality / Metrology
StatementInstrument passed all verification tests; certificate #CAL-2026-0188 issued, next verification 2027-06-25.